Manufacturer Correction – Philips SimplyGo Portable Oxygen Concentrators

By February 24, 2023March 6th, 2024Recalls Archive

About the Recall:

Philips Respironics has issued an urgent product notice regarding the Philips Respironics SimplyGo portable oxygen concentrator that could affect the performance of the equipment.

Philips identified a potential failure of SimplyGo Oxygen Concentrators built between July 27, 2021- January 18, 2022, if the concentrator is used on a continuous flow setting of 2 for 2,000 hours of operation, the caps could fail over time, resulting in an oxygen leak. When the oxygen leak begins, the compressor will compensate for the drop in pressure. However, once the crack propagates, a low O2 alarm will sound, and eventually the until will no longer function. It is possible that this failure will happen during the standard warranty period (2 years), depending on patient use.

Affected serial numbers are listed below:

Material Number: Serial Number

1068987: M32532944BE03
1068987: M3253305715E5
1068987: M32535448DC31
1068987: M325354986C49
1068987: M325355621A07
1068987: M3253565961D8
1068987: M32535757D27A
1068987: M3253587853FA
1068987: M32536008FDFD

If you have any questions or need help, please contact Philips customer service at 800-345-6443.