BACK TO RECALLS ARCHIVE
ResMed Magnet Masks- Urgent Field Notice
Nov 21, 2023
About the Notice:
ResMed is updating its existing contraindications and warnings to further inform patients and healthcare professionals on the safe use of ResMed masks with magnets. This is in response to new information obtained through post-market surveillance and industry practices related to possible magnetic interference when in close proximity to certain medical devices. Update to Contraindication In an update to the previously contraindicated metallic aneurysm hemostatic clips in the head and metallic splinters in one or both eyes following a penetrating eye injury, ResMed is further contraindicating the use of its masks with magnets for patients where they, or anyone in close physical contact while using the mask (e.g., bed partner), have the following:- Active medical implants that interact with magnets (i.e., pacemakers, implantable cardioverter defibrillators (ICD), neurostimulators, cerebrospinal fluid (CSF) shunts, insulin/infusion pumps)
- Metallic implants/objects containing ferromagnetic material (i.e., aneurysm clips/flow disruption devices, embolic coils, stents, valves, electrodes, implants to restore hearing or balance with implanted magnets, ocular implants, metallic splinters in the eye)
Product numbers/models that are affected:
| Item number | Item Description |
| 63331 | AirFit F30i Complete System - S/Std |
| 63333 | AirFit F30i Complete System - W/Std |
| 63330 | AirFit F30i Complete System - S/Sml |
| 63332 | AirFit F30i Complete System - M/Std |
| 63334 | AirFit F30i Complete System - M/LGE |
| 63401 | AirFit F20 System Medium w/hdgr |
| 64101 | AirFit F30 Complete Mask System: Medium |
| 63535 | AIRFIT N20 STARTER PK |
| 63501 | Airfit N20 System Medium w/hdgr |
| 63301 | AIRFIT F30I STD STPK - AMER |
| 63402 | Airfit F20 System Large w/hdgr |
| 63450 | AIRFIT F20 STARTER PACK |
Actions to be taken:
Contraindicated patients using a ResMed mask with magnets If you, or anyone in close physical contact while using the mask (e.g. bed partner),have an active medical implant or metallic implanted object that is identified within the Contraindications above:- Replace your mask containing magnets with an alternative mask without magnets in a timely manner. Contact your mask provider for alternative mask options.
- If an alternative mask is not available, consult your physician/doctor for appropriate next steps regarding your therapy.
- Dispose of the mask with magnets after you receive an alternative mask.
- If you are NOT contraindicated from using a ResMed mask with magnets, you may continue to use the mask, ensuring you follow all user instructions, including the updates described in this letter.
- Importantly, ResMed masks with magnets are to be kept at least 6 inches (150mm) away from implants or medical devices that may be adversely affected by magnetic interference, as described in the updated Warning.